Authors: Brett Lewis, Roland Teboh
Affiliation: HUMC, Hackensack University Medical Center
Abstract Preview: Purpose: The novel RefleXion SCINTIX BgRT system relies on the FDG uptake distribution from a PET study to plan and deliver dose. One of the patient eligibility criteria is that any uptake outside the...
Authors: Emily Hansen, Tyler Kaulfers, Alois M. Ndlovu, Francia Romero, Niara Rowe, Roland Teboh
Affiliation: Montefiore Einstein University Hospital, Hackensack University Medical Center
Abstract Preview: Purpose: Daily use and reliance on the auto registration feature requires a performance check. Clinical application of auto registration is performed without a selected region of interest (ROI), requi...
Authors: Resat Aydin, Brett Lewis, Jayesh Mistry, Roland Teboh
Affiliation: HUMC, Hackensack University Medical Center
Abstract Preview: Purpose:
We developed a Python-based solution to systematically quantify and detect fluorodeoxyglucose (FDG) uptake beyond a predefined Biologically Tracking Zone (BTZ) in PET images. This method i...
Authors: Emily Hansen, Jay Mistry, Alois M. Ndlovu, Niara Rowe, Roland Teboh
Affiliation: Hackensack University Medical Center
Abstract Preview: Purpose: SCINTIX Biology-guided radiotherapy (BgRT) on the RefleXion X1 relies on a new planning system and advanced planning techniques. The first patient SCINTIX BgRT plans are compared to similar c...
Authors: Resat Aydin, Joseph Barbiere, Brett Lewis, Roland Teboh
Affiliation: HUMC, Hackensack University Medical Center
Abstract Preview: Purpose:
Accurately compensating for respiratory-induced tumor motion is critical in BgRT, where precise delineation of volumes ensures effective dose delivery. We propose an integrated approach th...
Authors: Emily Hansen, Alois M. Ndlovu, Niara Rowe, Roland Teboh
Affiliation: Hackensack University Medical Center
Abstract Preview: Purpose: The accuracy of SCINTIX biology-guided radiotherapy (BgRT) on the RefleXion X1 is dependent on the correct alignment of the isocenter between all subsystems, including setup lasers, kV imagin...
Authors: Joseph Barbiere, Joseph Hanley, Brett Lewis, Jay Mistry, Alois M. Ndlovu, Roland Teboh
Affiliation: Hackensack University Medical Center, HUMC, Jersey Shore University Medical Center
Abstract Preview: Purpose:
Quantitative PET-CT can evaluate clinical response to radiotherapy. For tumors less than approximately 3cm partial volume effects due to the poor PET resolution makes accurate measurement ...
Authors: Emily Hansen, Tyler Kaulfers, Alois M. Ndlovu, Niara Rowe, Roland Teboh
Affiliation: Montefiore Einstein University Hospital, Hackensack University Medical Center
Abstract Preview: Purpose: A topic often discussed is the disparity of women in medical physics, overlooking the absence of Black and Hispanic women in the field. The intersectionality within marginalized communities i...