60RPM Rotational Winston Lutz Test to Characterize Operational MV-Kv Offset on Reflexion X1 System

Authors: Huixiao Chen, Ethan Hartowicz, Lichung Ku, Tim Perez, Ray Yang

Affiliation: Salem Health, RefleXion Medical, Department of Therapeutic Radiology, Yale University School of Medicine

Abstract Preview: Purpose: To characterize MV-kV isocenter offset on a novel ring gantry radiotherapy system where the Linac, PET, kV-CT, and MV imaging subsystems rotate together at native operating speed of 60RPM.

An Agglomerative Clustering-Based Program for Optimizing Multiple-Target SRS Treatment Planning

Authors: Josephine Chen, CheukKai Becket Hui, Yildirim D. Mutaf

Affiliation: Kaiser Permanente

Abstract Preview: Purpose:
To demonstrate the effectiveness of a target clustering program in generating cluster configurations and isocenter placements for multiple brain lesions in SRS treatment planning, with the...

Characterizing and Modeling Higher-Order Motions of Collimator Subsystem on a 60RPM Gantry Linac for Improved Small Target Dosimetry

Authors: David J. Carlson, Yiu-Hsin Chang, Huixiao Chen, Zhe (Jay) Chen, Blake Gaderlund, Ray Yang

Affiliation: RefleXion Medical, Department of Therapeutic Radiology, Yale University School of Medicine

Abstract Preview: Purpose: To address the root cause of dosimetric discrepancy for small central targets (<2cm) especially sensitive to geometric alignment of beamlets, by modeling higher-order oscillations of the coll...

Determining the Effect of Motion during Gamma Knife Stereotactic Radiosurgery on Anterior and Posterior Lesions Using an Anthropomorphic Phantom

Authors: Guang-Pei Chen, Nitish Chopra, Juan A. Garcia-Alvarez, Mi Huang, Slade J. Klawikowski, Haidy G. Nasief, George A. Noid, Abdul Parchur, Eric S. Paulson, Christina M. Sarosiek, Christopher Schultz

Affiliation: Department of Radiation Oncology, Medical College of Wisconsin

Abstract Preview: Purpose: During routine delivery of mask-based Gamma Knife (GK) SRS, motion is tracked via the position of a reflective sticker placed on the patient’s nose relative to two stationary reflectors on th...

Direct Table Rotational Isocenter Determination Using Radiation Isocenter

Authors: Christopher Ackerman, Chang Chang, Yansong Liu, Stephen Kenneth Northway, Bailey Vallejo

Affiliation: University of California San Diego / California Protons, California Protons Cancer Therapy Center, California Proton Cancer Therapy Center

Abstract Preview: Purpose: To determine mechanical isocenter of a 6-degree-of-freedom (6DOF) couch using radiation isocenter. Coincidence between the isocenters is quantified using MyQA Phoenix amorphous silicon detect...

Retrospective Analysis of Treatment Pauses Due to Intrafraction Movement for Frameless Single Fraction Gamma Knife Icon Radiosurgery

Authors: Nina Burbure, Tawfik G. Giaddui, Shidong Li, Curtis Miyamoto, Jeremy Price, Bin Wang

Affiliation: FCCC at Temple University Hospital

Abstract Preview: Purpose: Frameless gamma knife (GK) treatments use a thermoplastic mask coupled with high-definition motion management system that utilizes an infrared camera, and a reflective marker placed on the pa...

Uncertainty Analysis of Hyperarc VMAT Plans for Stereotactic Radiosurgery Patients with Multiple Brain Metastases

Authors: Shifeng Chen, Yannick P. Poirier, Huijun Xu, Byong Yong Yi, Baoshe Zhang, Hong Zhang, Jinghao Zhou

Affiliation: University of Maryland School of Medicine, University of Maryland Shore Regional Cancer Center, Vanderbilt University Medical Center, Department of Radiation Oncology, University of Maryland School of Medicine, Capital Region Medical Center

Abstract Preview: Purpose: To thoroughly assess the resilience of HyperArc VMAT plans for multiple brain metastases (MBT) in relation to (i) high-definition MLC leaf positional inaccuracies and (ii) patient rotational ...

Use of Radiopaque Rectal Spacer for IGRT – a Preliminary Analysis of Alignment Accuracy and Interobserver Variability

Authors: Joseph A. Miccio, Nicholas J. Potter, Zain Siddiqui

Affiliation: Penn State Health Milton S. Hershey Medical Center

Abstract Preview: Purpose: Fiducial markers are the gold standard for prostate alignment during CT-based image-guided radiotherapy (IGRT). Radiopaque hydrogel rectal spacers reduce rectal toxicity by displacing the rec...