Authors: Odunola Grace Babawale, B. Gino Fallone, Alireza Gazor, Andrei D. Ghila, Patricia A. K. Oliver, Michael W. Reynolds, Keith D. Wachowicz, Tania Rosalia Wood, Shima Y. Tari, Eugene Yip, Jihyun Yun
Affiliation: Nova Scotia Health, Dept. of Medical Physics and Dalhousie University, Dept. of Physics and Atmospheric Science, Dept. of Radiation Oncology, Dept. of Medical Physics, Cross Cancer Institute and Dept. of Oncology, University of Alberta; MagnetTx Oncology Solutions, www.magnetTX.com, Department of Medical Physics, Arthur J. E. Child Comprehensive Cancer Centre, Dept. of Medical Physics, Cross Cancer Institute and Dept. of Oncology, University of Alberta, Department of Medical Physics, BC Cancer, Medical Physics Division, Department of Oncology, University of Alberta and Department of Medical Physics, Cross Cancer Institute
Abstract Preview: Purpose: To evaluate the dose to a moving target and its surroundings in a magnetic field using 4D-Monte-Carlo (MC) simulation.
Methods: The study utilizes a previously validated MC model of our li...
Authors: William N. Duggar, Li Yuan
Affiliation: University of Mississippi Medical Center
Abstract Preview: Purpose:
Radiation Oncology departments typically utilize various systems from different vendors. Ensuring the integrity and correctness of data during transfers between these systems is essential ...
Authors: Laura Bennett, Wesley S. Culberson, Albert Du, Kevin J. Erhart, Ryan T. Flynn, Ryan Gardner, Alonso N. Gutierrez, Patrick M Hill, Daniel E. Hyer, Eric Jensen, Kaustubh A. Patwardhan, Blake R. Smith, Nhan Vu, Karsten K. Wake
Affiliation: University of Wisconsin, Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin - Madison, Miami Cancer Institute, Baptist Health South Florida, Department of Human Oncology, University of Wisconsin School of Medicine and Public Health, Ion Beam Applications (IBA), University of Iowa, Iowa Health Care, .decimal
Abstract Preview: Purpose: To develop and experimentally validate a novel Monte Carlo (MC) framework designed to enable calculation-based quality assurance (QA) for collimated intensity modulated proton therapy treatme...
Authors: Christopher J. Mart, Inna V. McGowin
Affiliation: Medical University of South Carolina
Abstract Preview: Purpose: Electron treatments are prone to setup variations and uncertainties due to absence of imaging or surface guided alignment. Patient setup relies on daily marking and visual verification. A 3D-...
Authors: Raneem Atta, Alejandro Bertolet, Mislav Bobić, Wesley E. Bolch, Robert Joseph Dawson, Carlos Huesa-Berral, Harald Paganetti, Eric Wehrenberg-Klee
Affiliation: Massachusetts General Hospital, University of Florida, Department of Radiation Oncology, Massachusetts General Hospital and Harvard Medical School, Massachusetts General Hospital and Harvard Medical School, Department of Radiology, Massachusetts General Hospital and Harvard Medical School
Abstract Preview: Purpose: Representations of intra-organ vasculature have a variety of uses in the field of computational dosimetry but generally rely on models derived from population-averaged reference individuals. ...
Authors: Denisa R. Goia, M. Saiful Huq, Ronald John Lalonde, Fang Li, Noor Mail, Joseph Shields, Christopher Tyerech
Affiliation: UPMC Hillman Cancer Center, UPMC Hillman Cancer Center and University of Pittsburgh School of Medicine, Department of Radiation Oncology, University of Pennsylvania, UPMC
Abstract Preview: Purpose: HyperSight is a new platform for image-guided radiation therapy, offering advanced reconstruction algorithms, a large field-of-view, and rapid acquisition times. To validate the performance o...
Authors: Juan-Francisco Calvo-Ortega, Andrew Cousins, Ashley Cullen, Andrew Dipulia, Peter B. Greer, Seng Boh Gary Lim, Shih-Chi Lin, D. Michael Lovelock, Conor McGarry, Victoria Robinson, Cameron Stanton, Baozhou Sun, Ching-Ling Teng, Gemma Warner, Benjamin J. Zwan
Affiliation: Northern Ireland Cancer Centre, Baylor College of Medicine, Department of Medical Physics, Memorial Sloan Kettering Cancer Center, Christchurch Hospital, Northwell, Central Coast Cancer Centre, Calvary Mater Hospital, Hospital Quironsalud Barcelona, Icahn School of Medicine at Mount Sinai, Chris O'Brien Lifehouse, University of Newcastle
Abstract Preview: Purpose: This multi-institution Electronic Silicon-based Remote Survey of Small-field Output (ESPRESSO) study aims to develop the remote audit process to evaluate the safety of single-isocenter multi-...
Authors: Debora de Souza Antonio, Romy Guthier, Konrad Pawel Nesteruk, Erno Sajo, William Paul Segars, Gregory C. Sharp, Atchar Sudhyadhom, Hengyong Yu
Affiliation: Massachusetts General Hospital, Carl E. Ravin Advanced Imaging Laboratories and Center for Virtual Imaging Trials, Duke University Medical Center, Brigham and Women’s Hospital and Dana Farber Cancer Institute, Harvard Medical School,, Massachusetts General Hospital and Harvard Medical School, University of Massachusetts Lowell
Abstract Preview: Purpose: To develop an open-access toolkit for rapidly generating simultaneously realistic CT scans and low-field MR images of the abdominal region, based on patient data, while employing an XCAT phan...
Authors: Joyoni Dey, Hunter Cole Meyer, Murtuza Taqi
Affiliation: Louisiana State University
Abstract Preview: Purpose: We propose using super-resolution methods for X-ray grating interferometry without an analyzer with detectors that fail to meet the Nyquist sampling rate needed for traditional image recovery...
Authors: Javier Chiasson, Dominique Guillet, Arthur Lalonde, Karim Zerouali
Affiliation: Universite de Montreal, Centre Hospitalier de l'Universite de Montreal
Abstract Preview: Purpose:
To validate a Monte Carlo (MC) model of a Ultra-High dose rate (UHDR) electron beam using a prototype flashdiamond (fD) detector coupled to an electrostatically shielded beam current trans...
Authors: Brianna Baek, David J. Carlson, Zhe (Jay) Chen, Dae Yup Han, Bon Tack Koo, Hyojun Park
Affiliation: Columbia University, Canyon Crest Academy, Department of Therapeutic Radiology, Yale University School of Medicine, Department of Radiation Oncology, Chung-Ang University Gwang Myeong Hospital
Abstract Preview: Purpose: To find an optimized treatment technique and shielding design for minimizing the dose to a fetus, a Monte-Carlo simulation was performed to evaluate the dosimetric contribution.
Methods: G...
Authors: Lily Jo Bertemes, Careesa Billante, Ashley Cetnar, Maximilian Stephen Meineke, Runhe Tan
Affiliation: The James Cancer Center, The Ohio State University, The Ohio State University - James Cancer Hospital
Abstract Preview: Purpose: The response of healthy vasculature due to radiation treatments is an underexplored area of radiation oncology. Understanding the impact of radiation on veins could advance treatment optimiza...
Authors: Hongyu Jiang, Wangyao Li, Dima Soultan, Fen Wang, Jun Xu
Affiliation: Department of Radiation Oncology, University of Kansas Medical Center
Abstract Preview: Purpose: To validate the Auto-Beam Hold functionality via the MMI connection between Varian SGRT system IDENTIFY 3.0 and treatment linacs.
Methods: A treatment plan was created in the Eclipse TPS u...
Authors: Mehdi Amini, Minerva Becker, Simina Chiriac, Alexandre Cusin, Dimitrios Daskalou, Ghasem Hajianfar, Sophie Neveu, Marcella Pucci, Yazdan Salimi, Pascal Senn, Habib Zaidi
Affiliation: Geneva University Hospital, Division of Radiology, Diagnostic Department, Geneva University Hospitals, Service of Otorhinolaryngology-Head and Neck Surgery, Department of Clinical Neurosciences, Geneva University Hospitals
Abstract Preview: Purpose: Personalized prediction of vestibular schwannoma (VS) tumour growth is crucial for guiding patient management decisions toward observation versus intervention. This study proposes an automate...
Authors: Yankun Lang, Lei Ren, Dario B. Rodrigues
Affiliation: University of Maryland School of Medicine, Department of Radiation Oncology, University of Maryland School of Medicine
Abstract Preview: Purpose:
HTP of microwave (MW) phased-array systems determine MW antenna settings to maximize energy absorption (SAR in W/kg) in tumor. Conventional HTP algorithms calculate SAR based on electromag...
Authors: Michael Ashenafi, Mario Ramos Gallardo, Amy Herman, Sean M. Tanny
Affiliation: Department of Radiation Oncology, University of Rochester
Abstract Preview: Purpose: Updates to the Varian Eclipse Treatment Planning System (TPS) have introduced new objects to replace pre-existing Clinical Protocols to aide clinicians in plan quality assessment. We have cre...
Authors: Craig K Abbey, John M. Boone, Richard E. Colbeth, Andrew M. Hernandez, Vance Robinson, Paul Schwoebel, Jeffrey H. Siewerdsen, Alejandro Sisniega, Wojciech B. Zbijewski, Huanyi Zhou
Affiliation: UC Santa Barbara, University of California, UT MD Anderson Cancer Center, Johns Hopkins University, University of New Mexico Albuquerque, UC Davis Health, Varex Imaging Corporation
Abstract Preview: Purpose: A new CT configuration utilizing a multiple X-ray-source array (MXA) is under development to mitigate cone-beam (CB) artifacts in single-rotation wide coverage imaging. We investigate the sam...
Authors: Hua-Chieh Shao, Chenyang Shen, Jiacheng Xie, Shunyu Yan, You Zhang
Affiliation: Medical Artificial Intelligence and Automation (MAIA) Lab, Department of Radiation Oncology, UT Southwestern Medical Center, UT Southwestern Medical Center, Medical Artificial Intelligence and Automation (MAIA) Laboratory, Department of Radiation Oncology, UT Southwestern Medical Center, Medical Artificial Intelligence and Automation (MAIA) Laboratory, Department of Radiation Oncology, UT Southwestern Medical Center, The University of Texas at Dallas
Abstract Preview: Purpose: Motion-resolved CBCT imaging, which reconstructs a dynamic sequence of CBCTs reflecting intra-scan motion (one CBCT per x-ray projection), is highly desired for regular/irregular motion chara...
Authors: Peter Balter, Kristy K. Brock, Clifton David Fuller, Tze Yee Lim, James Long, Brigid A. McDonald, Androniki Mitrou, Andrew J. Schaefer, Stina Svensson
Affiliation: The University of Texas MD Anderson Cancer Center, RaySearch Laboratories, Rice University, Department of Radiation Oncology, The University of Texas MD Anderson Cancer Center, The University of Texas MD Anderson Cancer Center UTHealth Houston Graduate School of Biomedical Sciences
Abstract Preview: Purpose: CBCT-based accumulated dose can provide up-to-date dose information on responding anatomy. To optimize resources, it is necessary to determine how many scans provide adequate information for ...
Authors: Liyuan Chen, Steve Jiang, Chenyang Shen
Affiliation: Department of Radiation Oncology, UT Southwestern Medical Center, UT Southwestern Medical Center
Abstract Preview: Purpose: Delays in radiation therapy (RT) initiation caused by conventional CT simulation processes can hinder timely treatment delivery and patient outcomes. This study proposes a Virtual Treatment S...
Authors: Ryan Clark, Anthony Magliari, Luis Felipe Oliveira E Silva, Yang Sheng, Jason R Vickress, Qingrong Jackie Wu, Giulianne Rivelli Rodrigues Zaratim
Affiliation: Verspeeten Family Cancer Center, Office of Medical Affairs, Varian, A Siemens Healthineers Company, Confiar Radioterapia, Duke University Medical Center
Abstract Preview: Purpose: Develop a toolset to replace several of the tedious steps required to manually generate 3D breast treatment plans using static gantry forward planned tangent beams.
Methods: Eclipse Sc...
Authors: Sam Beddar, Jason Michael Holmes, Daniel G. Robertson, James J. Sohn, Ethan D. Stolen
Affiliation: Department of Radiation Oncology, Mayo Clinic, MD Anderson Cancer Center, Department of Radiation & Cellular Oncology, University of Chicago, Department of Radiation and Cellular Oncology, University of Chicago
Abstract Preview: Purpose: Camera-based scintillation dosimetry incorporating large volumes have shown promise for fast and comprehensive evaluation of external beam treatment fields. While some efforts have been made ...
Authors: Arjit K. Baghwala, Sunan Cui, Jessica Fagerstrom, Eric C. Ford, Kristi Rae Gayle Hendrickson, Sharareh Koufigar, Samuel Ming Ho Luk, Bishwambhar Sengupta, Afua A. Yorke
Affiliation: University of Washington, Department of Radiation Oncology, Fred Hutchinson Cancer Center, University of Washington, Department of Radiation Oncology, University of Vermont Medical Center, University of Washington and Fred Hutchinson Cancer Center, Houston Methodist Hospital
Abstract Preview: Purpose: The global burden of cancer continues to rise, leading to an increased workload in radiation oncology clinics. This surge is not only due to the growing demand for treatment machines and moda...
Authors: Mateb Ali Alghamwa, Haya Aljuaid, Siyong Kim, Tianjun Ma, William Song
Affiliation: Virginia Commonwealth University
Abstract Preview: Purpose: The integration of Plastic Scintillation Detectors (PSDs) into 0.35T MR-Linac systems represents a significant advancement in radiotherapy, offering substantial benefits due to their tissue-e...
Authors: Kaile Li, Esmail Parsai
Affiliation: Alexander T. Augusta Military Medical Center, Alexander T. Augusta Miliary Medical Center
Abstract Preview: Purpose: To evaluate the sensitivity of the portal dosimetry system with the Arc CHECK system at identical Volumetric Modulated Arc Therapy (VMAT) setup
Methods: An Arc CHECK system from Sun Nuclea...
Authors: Mark E Artz, Jiyeon Park, Lori Rice, Mohammad Saki, Sanjeev Shukla, Michael Vieceli
Affiliation: University of Florida Health Proton Therapy Institute, Department of Radiation Oncology, University of Florida College of Medicine, UF Health Proton Therapy Institute, Department of Urology, University of Florida College of Medicine
Abstract Preview: Purpose: To investigate how varying levels of DNA repair mutations affect the LET sensitivity of prostate cancer cells.
Methods: The prostate cancer cell lines Du145 (low DNA repair mutations), PC3...
Authors: Xiaoda Cong, Rohan Deraniyagala, Xuanfeng Ding, Xiaoqiang Li, Jian Liang, Peilin Liu, Craig Stevens, Xiangkun Xu, Weili Zheng
Affiliation: Corewell Health William Beaumont University Hospital, Corewellhealth William Beaumont University Hospital, William Beaumont University Hospital, Corewellhealth William Beaumont Hospital, Department of Radiation Oncology, Corewell Health William Beaumont University Hospital
Abstract Preview: Purpose:
Commission a step-and-shoot arc therapy(SPArc-step&shoot) for treating head-neck cancer patients as a desired interim milestone toward full dynamic treatment.
Methods:
An in-house de...
Authors: Niko Ahjoniemi, Antti Kulmala, Katja Peltola
Affiliation: Clinical Research Institute HUS, Helsinki University Hospital
Abstract Preview: Purpose: To determine the backlash of a multileaf collimator (MLC) using four measurement methods: motor encoder (primary readout), linear potentiometer (secondary readout), amorphous silicon panel (p...
Authors: Junsheng Cao, Khushdeep Singh
Affiliation: Overlook Medical Center
Abstract Preview: Purpose: Eclipse v18 introduces Enhanced Leaf Modeling for MLCs. This study investigates the MLC off-axis characteristics using portal dosimetry
Methods:
MLC characteristics were analyzed for th...
Authors: Tetsuo Fukuda, Hideaki Hirashima, Kohei Kawata, Yukako Kishigami, Takashi Mizowaki, Mitsuhiro Nakamura, Yohei Sawada, Maika Urago
Affiliation: Kyoto University, kyotoUnivercity, Hitachi high-tech, Department of Radiation Oncology and Image-applied Therapy, Graduate School of Medicine, Kyoto University
Abstract Preview: Purpose: To present the comprehensive commissioning results of the O-ring shaped linac, OXRAY.
Materials and Methods: OXRAY enables a novel three-dimensional unicursal noncoplanar irradiation metho...
Authors: Sorour Hosseini, YuHuei Jessica Huang, Jeremy Kunz, Nicholas Pierre Nelson, Ryan G. Price, Prema Rassiah, Hui Zhao
Affiliation: University of Utah, Department of Radiation Oncology, Huntsman Cancer Institute, University of Utah
Abstract Preview: Purpose: To investigate the feasibility and clinical implications of creating customized workflows in a new SGRT system: LUNA 3D (LAP, Florida).
Methods: Three clinical SGRT workflows were designed...
Authors: Carri K. Glide-Hurst, Thomas M Grist, Kevin M. Johnson, Prashant Nagpal, Tarun Naren, Chase Ruff, Oliver Wieben, Jiwei Zhao
Affiliation: Department of Biostatistics and Medical Informatics, University of Wisconsin - Madison, Department of Radiology, University of Wisconsin-Madison, Department of Medical Physics, University of Wisconsin-Madison, Departments of Human Oncology and Medical Physics, University of Wisconsin-Madison
Abstract Preview: Purpose: Cardiotoxicity is a devastating side effect for thoracic radiotherapy (RT). Currently, standard RT imaging is insufficient to decouple cardiorespiratory motion, limiting substructure-specific...
Authors: Ting Chen, Siming Lu, Tanxia Qu, Hesheng Wang
Affiliation: NYU Langone Health
Abstract Preview: Purpose: This study, the first in a two-part series, aims to determine the isocenter position and size for the MR Linac UNITY, focusing on the central axis (CAX) alignment of its Multi-Leaf Collimator...
Authors: Ashley Elizabeth Bonner, Michael G. Snyder
Affiliation: Corewell Health William Beaumont University Hospital, Department of Radiation Oncology, Corewell Health William Beaumont University Hospital
Abstract Preview: Purpose: Develop and evaluate metrics for assessing beam matching of profiles across Elekta linacs in a multisite clinical setting.
Methods: Four Elekta linear accelerators (three VersaHD and one A...
Authors: Wesley E. Bolch, Lotem Buchbinder Shadur, Natalia Estefania Carrasco-Rojas, Chansoo Choi, Robert Joseph Dawson, Shaheen Dewji, John Francis, Emmanual Mate-Kole, Wyatt Smither, Yitian Wang
Affiliation: Georgia Tech, University of Florida, Nuclear and Radiological Engineering and Medical Physics Programs, Georgia Institute of Technology
Abstract Preview: Purpose: Develop a software application capable of assessing the radiation contamination of military service members and members of the general public for fast, on-site radiological triage. The softwa...
Authors: Yona Choi, Kyo-tai Kim, Seungwoo Park, Dung Thi Tran
Affiliation: Gemss Health Care Inc,, Korea Institute of Radiological & Medical Sciences
Abstract Preview: Purpose: We developed shielding materials and a hybrid concept collimator for fluoroscopy to reduce the dose to normal tissues outside the treatment field, which are capable of quantifying dose rate a...
Authors: Adam P. Berdusco, Matthew R. Ceelen, Renata Farrell, Carri K. Glide-Hurst, Will Martin, Charles F. Maysack-Landry, Morgan A. McGauley, James Rice, Jordan M. Slagowski, Yuhao Yan
Affiliation: Departments of Human Oncology and Medical Physics, University of Wisconsin-Madison, University of Wisconsin - Madison, Department of Human Oncology, University of Wisconsin-Madison
Abstract Preview: Purpose: Upright radiotherapy (RT) is now available although this novel vertical CT and positioner presents unique challenges for quality assurance and testing methodologies. We have designed and buil...
Authors: Nellie Brovold, Wenbo Gu, Huan Amanda Liu, Lin Ma
Affiliation: University of Pennsylvania, Mayo Clinic
Abstract Preview: Purpose: To develop a low-cost comprehensive QA phantom and software for single-iso multi-target SRS Linac machines. This phantom can perform 1) Winston-Lutz QA test for on-axis and off-axis targets, ...
Authors: Huiming Dong, Jonathan Pham, X. Sharon Qi, Ann Raldow
Affiliation: Department of Radiation Oncology, University of California, Los Angeles
Abstract Preview: Purpose: The study aimed to investigate longitudinal apparent diffusion coefficient (ADC) as an early biomarker of treatment response in patients with locally advanced rectal cancer (LARC) undergoing ...
Authors: Ian J. Butterwick, Sumudu Katugampola, ChangSeon Kim, Evan G. Meekins, Zhengdong Zhang
Affiliation: Mount Sinai Beth Israel, Gesinger Medical Center, Geisinger Medical Center
Abstract Preview: Purpose: To evaluate the dosimetric impact of enhanced dynamic leaf gap (EDLG) modeling in Eclipse V18.0 on small-field dosimetry within VMAT plans.
Methods: Plans with small fields (1 cm x 1 cm to...
Authors: Enzo Charles Aucca, Roger Challco, Andres M. Gonzales Galvez, Carmen SANDRA Guzman Calcina, Jimmy Hernandez-Bello, Modesto Montoya, Renzo Aaron Ocampo, Erick Paniagua, Miguel Alberto Risco-Castillo, Dante E. Roa, Renzo E. Romero, William Souza Santos, Mirko Salomón Alva Sánchez
Affiliation: Graduate Program in Information Technologies and Health Management, University of Porto AlegreFederal University of Health Sciences of Porto Alegre, Universidad Tecnologica del Peru, Pontificia Universidad Catolica del Peru, Electronic Engineering Department, Leonard Cancer Institute - Radiation Oncology, Universidad Nacional de Ingenieria, Universidad Nacional de Ingenieria, Electronic Engineering Department, Federal University of Sergipe, HRS Oncology International, Universidad Nacional Federico Villareal
Abstract Preview: Purpose: Provide Monte Carlo (MC) and experimental dosimetric assessments of a 3D cubic electronic detector array based on percent depth doses (PDDs), profiles, linearity, and dose rate response of ph...
Authors: Kazi Towmim Afrin, Imad M. Ali, Jordan West
Affiliation: University of Oklahoma Health Sciences Center, University of Oklahoma
Abstract Preview: Purpose: To evaluate quantitatively the dosimetric uncertainties associated with dose calculation and measurement of small photon fields used in stereotactic photon therapy.
Methods: The output fac...
Authors: J Isabelle Choi, Meng Wei Ho, Isabel Lin, Hang Qi, Qi Zeng, Ajay Zheng
Affiliation: New York Proton Center, Scarsdale High School
Abstract Preview: Purpose: Proton therapy is a promising treatment modality for breast cancer due to its unique dosimetric advantage. Accurate dose calculation is critical to ensure effective treatment outcomes. The fo...
Authors: Lian Duan, Stephen F. Kry, Hunter S. Mehrens, Christine Peterson, Paige A. Taylor
Affiliation: The University of Texas MD Anderson Cancer Center, UT MD Anderson Cancer Center
Abstract Preview: Purpose: To develop predictive models for IROC SRS head phantom audits and to identify important factors influencing institutional performance.
Methods: The IROC SRS head phantom includes two TLDs ...
Authors: Yang Lei, Haibo Lin, Tian Liu, Charles B. Simone, Shouyi Wei, Ajay Zheng
Affiliation: Icahn School of Medicine at Mount Sinai, New York Proton Center
Abstract Preview: Purpose: Radiation-induced lung injury (RILI), encompassing pneumonitis and fibrosis, represents a critical dose-limiting factor in lung cancer radiation therapy. Variability in treatment outcomes is ...
Authors: Nema Bassiri, Kristen A. Duke, Neil A. Kirby, Thomas Wilson Klender, Christien Kluwe, Niko Papanikolaou, Holly Paschal, Karl H. Rasmussen, Daniel L. Saenz, Michelle de Oliveira
Affiliation: Miami Cancer Institute, University of Texas HSC San Antonio, UT Health San Antonio
Abstract Preview: Purpose: Determine the practical usability for three phantoms to be used with the ExacTrac Dynamic thermal imaging system. Thermal-guided imaging can be used for inter and intra motion management, pat...
Authors: Blessing Chinelo Akah, Daniel W. Neck, David J. Perrin, Sotirios Stathakis
Affiliation: Mary Bird Perkins Cancer Center
Abstract Preview: Purpose: To assess the reliability of the Aktina Isopoint software in evaluating different components of the linear accelerator for optimized quality assurance and linac performance
Methods: The Ak...
Authors: Sunil Mani, Ilakiya Raghavendiran, Vasanthan Sakthivel, Skanda Kabilan Swamy
Affiliation: Edmond Santa Fe High School, Edmnd Deer Creek High, Cancer Specialists of North Florida, Advanced Medical Physics
Abstract Preview: Purpose: Purpose/Objective(s): To evaluate and validate the dose calculation and delivery accuracy for patients planned using diagnostic CT
Methods: A total of forty-two patients with palliative sp...
Authors: Shatha Al Afifi, Hana Baroudi, Leonard Che Fru, Laurence Edward Court, Suzanne B. Evans, Kent A. Gifford, Adam D. Melancon, Melissa P. Mitchell, Issa Mohamad, Patricia Murina, Manickam Muruganandham, Tucker J. Netherton, Callistus M. Nguyen, Joshua S. Niedzielski, Deborah L. Schofield, Simona Shaitelman, Willie Shaw, Sanjay S. Shete, Adam Shulman, Brendon Smith, Sheeba Thengumpallil, Carlos Daniel Venencia, Conny Vrieling
Affiliation: University of Cape Town, MD Anderson Cancer Center, The University of Texas MD Anderson Cancer Center, University of the Free State, UT MD Anderson Cancer Center, King Hussein Cancer Center, Instituto Zunino - Fundacion Marie Curie, Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Dra., Hirslanden Clinique des Grangettes, Department of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Department of Therapeutic Radiology, Yale University School of Medicine, Yale University, The University of Texas, MD Anderson, Houston TX
Abstract Preview: Purpose:
Automated contouring and planning tools are usually trained on single-institution datasets, seldom tested across diverse patient populations. This introduces a risk of population bias, res...
Authors: Weixing Cai, Laura I. Cervino, Yabo Fu, Xiuxiu He, Tianfang Li, Xiang Li, Hao Zhang
Affiliation: Department of Medical Physics, Memorial Sloan Kettering Cancer Center
Abstract Preview: Purpose:
This work aims to develop an innovative technique to evaluate patients’ daily respiratory pattern using three-dimensional (3D) deformation vector fields (DVF) derived from a free-breathing...
Authors: Chunhui Han, An Liu, Anthony Magliari, Lesley Rosa
Affiliation: Office of Medical Affairs, Varian, A Siemens Healthineers Company, Department of Radiation Oncology, City of Hope National Medical Center
Abstract Preview: Purpose: To evaluate the performance of a knowledge-based planning (KBP) optimization model for total marrow and lymphoid irradiation (TMLI) on a ring-gantry linac with a dual-layer MLC and to evaluat...
Authors: Bing-Hao Chiang, Eric C. Ford, Juergen Meyer, Timothy D. Solberg
Affiliation: Department of Radiation Oncology, Fred Hutchinson Cancer Center, University of Washington
Abstract Preview: Purpose: This study evaluated the potential risk reduction of various pretreatment patient-specific quality assurance (PSQA) programs for Intensity Modulated Radiation Therapy (IMRT) in mitigating fai...
Authors: Min Cheol Han, Hojin Kim, Jin Sung Kim, Hyejin Lee, Sac Lee, Yongdo Yun
Affiliation: Department of Radiation Oncology, Yonsei Cancer Center, Heavy Ion Therapy Research Institute, Yonsei University College of Medicine
Abstract Preview: Purpose: Carbon ion radiotherapy (CIRT) is characterized by high relative biological effectiveness (RBE), which must be integrated into the treatment planning process. This study aims to develop a nov...
Authors: Yoshihiko Hoshino, Kenji Hotta, Taeko Matsuura, Ai Nakaoka, Hidenobu Tachibana
Affiliation: Hokkaido University, Department of Radiology, Gunma University Hospital, Radiation Safety and Quality Assurance Division, National Cancer Center Hospital East
Abstract Preview: Purpose: Gel dosimeters enable three-dimensional dose measurements, but no reports have evaluated the fundamental performance of VIPER-type gel dosimeters in proton beam measurements. Therefore, we ev...
Authors: Resat Aydin, Joseph Barbiere, Brett Lewis, Roland Teboh
Affiliation: HUMC, Hackensack University Medical Center
Abstract Preview: Purpose:
Accurately compensating for respiratory-induced tumor motion is critical in BgRT, where precise delineation of volumes ensures effective dose delivery. We propose an integrated approach th...
Authors: Uriel Aura, Qing Chen, Arpit M. Chhabra, J Isabelle Choi, Chanda Guha, Meng Wei Ho, Sheng Huang, Minglei Kang, Stanislav Lazarev, Nancy Y Lee, Yang Lei, Haibo Lin, Hang Qi, Mahbubur Rahman, Charles B. Simone, Shouyi Wei, Irini Yacoub, Francis Yu, Anna Zhai, Ajay Zheng
Affiliation: DYN'R Medical System, New York Proton Center, Montefiore Medical Center and Albert Einstein College of Medicine, Icahn School of Medicine at Mount Sinai, Department of Radiation Oncology, Memorial Sloan Kettering Cancer Center, Tianjin Medical University Cancer Institute&Hospital, Department of Human Oncology, University of Wisconsin-Madison
Abstract Preview: Purpose: Managing tumor motion during pencil beam scanning proton therapy remains challenging. This study investigated the feasibility and dosimetric benefits of Free-Breathing Respiratory-Gated Deliv...
Authors: Matthew R. Hoerner, Maryam Naseri, Mena Shenouda
Affiliation: Yale University School of Medicine, Yale University
Abstract Preview: Purpose:
Historically film has been the gold standard receptor for measurements of x-ray tube focal spot size. Flat Panel Digital Radiography Detectors (FPDRDs) have lower spatial resolution compar...
Authors: Rodrigo T Massera, Sofia Giaccone Thomaz, Alessandra Tomal, Giovanna Tramontin
Affiliation: Universidade Estadual de Campinas. Instituto de Física Gleb Wataghin, Department of Imaging & Pathology, unit of Medical Physics & Quality Assessment, KU Leuven
Abstract Preview: Purpose: Monte Carlo simulations are increasingly used in breast dosimetry for their precision in estimating difficult-to-measure quantities, such as glandular dose. With ionizing radiation in breast ...
Authors: Caroline Esposito, Keith T Griffin, Jae Won Jung, Choonik Lee, Choonsik Lee, Matthew Mille, Harald Paganetti, Sergio Morato Rafet, Jan PO Schuemann, Jungwook Shin, Torunn I Yock
Affiliation: East Carolina University, University of Michigan, Massachusetts General Hospital, National Cancer Institute, Division of Cancer Epidemiology and Genetics, National Cancer Institute, Massachusetts General Hospital and Harvard Medical School
Abstract Preview: Purpose: The National Cancer Institute’s Pediatric Proton and Photon Therapy Comparison Cohort aims to collect and analyze data from cancer centers across the United States and Canada to quantify diff...
Authors: Min-Sig Hwang, Danny K. Lee, Daniel C. Pavord, Kyung Lim Yun
Affiliation: Allegheny Health Network
Abstract Preview: Purpose: Ensuring the quality of treatment plans through patient-specific pre-treatment quality assurance (QA) is essential. However, the use of physical phantom-based QA devices is not feasible for o...
Authors: Youssef Ben Bouchta, Chen Cheng, Owen Thomas Dillon, Mark Gardner, Paul J. Keall, Purnima Sundaresan
Affiliation: Radiation Oncology Network, Western Sydney Local Health District, Image X Institute, Faculty of Medicine and Health, The University of Sydney
Abstract Preview: Purpose: There are three clinical motivations for real time IGRT in head and neck cancer radiation therapy: (1)50% of patients experience anxiety from the patient mask (2)patient motion still occurs d...
Authors: Clifford Eskey, Scott Gordon-Wylie, Scott Gordon-Wylie, Dhrubo Jyoti, Daniel Reeves, John B. Weaver
Affiliation: Flow Aluminum, Dartmouth College, Fred Hutch Cancer Center, University of Washington, Dartmouth-Health
Abstract Preview: Purpose: Mechanical thrombectomy (MT) is the preferred treatment for ischemic stroke from large vessel occlusion. The thrombus or clot is extracted from the vessel via a catheter using a stent retriev...
Authors: Vibha Chaswal, Stephen D. Davis, Alonso N. Gutierrez, Noah S. Kalman, Yongsook C. Lee, William Romaguera, Ranjini P. Tolakanahalli, D. Jay J. Wieczorek
Affiliation: Miami Cancer Institute, Baptist Health South Florida
Abstract Preview: Purpose: To report machine performance of our superficial radiation therapy unit (SRT-100 Vision, Sensus Healthcare, Boca Raton, FL) for the first three years after commissioning of the unit.
Metho...
Authors: Evan Barber, Gloria P. Beyer, Matthew J. Daniels, Callum Hartley, Jordi Saez, Philip Wheeler
Affiliation: TrueNorth Medical Physics, Department of Radiotherapy Physics, Velindre University NHS Trust, Department of Radiation Oncology, Hospital Clinic de Barcelona, Medical Physics Services, LLC
Abstract Preview: Purpose: To quantify inter-system and inter-layer dosimetric leaf gap (DLG) variations in Halcyon/Ethos multileaf collimators (MLCs) using custom sweeping gap tests, and to evaluate implications for c...
Authors: Christopher S. Melhus, Elizabeth Meyer, Cassandra Stambaugh
Affiliation: Tufts Medical Center
Abstract Preview: Purpose: To investigate the clinical treatment planning parameters that impact patient-specific quality assurance (PSQA) passing rates for hippocampal avoidance whole brain (HAWB) treatment plans.
...
Authors: Tess Armstrong, Nema Bassiri, Alonso N. Gutierrez, Michael Kasper, Natalia Lutsik, Eric Mellon, Kathryn E. Mittauer, Siamak P. Nejad-Davarani, Shyam Pokharel, Suresh Rana, Hui Wang, Joseph Weygand
Affiliation: Department of Radiation Oncology and Applied Science, Dartmouth Health, Miami Cancer Institute, Baptist Health South Florida, ViewRay, Inc., Miami Cancer Institute, Lynn Cancer Institute, Boca Raton Regional Hospital, Baptist Health South Florida, Department of Radiation Oncology, University of Miami
Abstract Preview: Purpose: Radiation treatments on the MR-linac (MRL) enable daily acquisition of anatomical and physiological images for adaptive treatment planning. The apparent diffusion coefficient (ADC) estimated ...
Authors: Jiayi Chen, Shaolei Li, Fuhua Yan, Yingli Yang, Jie Zhang
Affiliation: Department of Radiology, Ruijin Hospital, Institute for Medical Imaging Technology, Ruijin Hospital, Department of Radiology, Ruijin Hospital Shanghai Jiaotong University School of Medicine, Shanghai United imaging Healthcare Advanced Technology Research Institute, Department of Radiation Oncology, Ruijin Hospital
Abstract Preview: Purpose: The aim of this exploratory study is to investigate the feasibility of establishing a model to explore tissue component heterogeneity using multidimensional diffusion magnetic resonance imagi...
Authors: Jeff Boob, Jaydev K. Dave, Tim Minch, Bhavyata Tanna
Affiliation: Dr. Ronald E. McNair Academic High School, Mayo Clinic, CurveBeam AI
Abstract Preview: Purpose: To estimate effective doses associated with different imaging protocols for hip imaging with a cone beam CT (CBCT) scanner specifically designed for weight-bearing acquisitions.
Methods: A...
Authors: Panagiotis Iliopoulos, Marios Myronakis, Kyriaki Theodorou
Affiliation: Medical Physics Department, Medical School, University of Thessaly
Abstract Preview: Purpose: To estimate individualized absorbed dose at Organs at Risk (OARs) from kV Cone Beam CT (CBCT) imaging in the pelvic area, prior radiotherapy, and assess excess risk for secondary malignancies...
Authors: Issra Rashed, Jennie Wilder, Jingeng Zhu
Affiliation: Rush Riverside Cancer Institute, Advanced Oncology Solutions (AOS) Varian, a Siemens Healthineers company
Abstract Preview: Purpose: Develop a method to plan and deliver homogeneous dose to Craniospinal(CS) irradiation on Linear Accelerator.
Methods: Three treatment planning volumes (PTV) are created in sequence from su...
Authors: Ning Wen, Zheyu Zhang
Affiliation: Department of Radiology, Ruijin Hospital Shanghai Jiaotong University School of Medicine, Shanghai Jiaotong University
Abstract Preview: Purpose: The graduate course, “Principles of Medical Imaging,” aims to advance imaging technology by integrating artificial intelligence (AI) into medical imaging. It bridges interdisciplinary fields,...
Authors: Richard Bakst, Arpit M. Chhabra, J Isabelle Choi, Chanda Guha, Minglei Kang, Nancy Y Lee, Haibo Lin, Hang Qi, Charles B. Simone, Pingfang Tsai, Milo Vermeulen, Shouyi Wei, Xiaodong Wu, Lee Xu, Irini Yacoub, Xingyi Zhao, Ajay Zheng
Affiliation: University of Miami, Peking University, New York Proton Center, Montefiore Medical Center and Albert Einstein College of Medicine, icahn school of medicine at mount sinai, Department of Radiation Oncology, Memorial Sloan Kettering Cancer Center, Department of Human Oncology, University of Wisconsin-Madison
Abstract Preview: Purpose:
Three-dimensional lattice radiation therapy (3D-LRT) effectively achieves local tumor control with limited morbidity. Pencil beam scanning (PBS) proton therapy provides precise dose confor...
Authors: Gueorgui Gueorguiev, Kavya Thaniyarasu
Affiliation: University of Bologna, National Centre of Radiobiology and Radiation Protection
Abstract Preview: Purpose: This study investigates the clinical viability of double scattering proton therapy treatment planning for thoracic malignancies using single or multiple treatment fields. The aim of this stud...
Authors: Zakaria Aboulbanine, Greeshma A. Agasthya, Paul Inman, Anuj J. Kapadia, Anthony Hong Cheol Lim, Jayasai Ram Rajagopal
Affiliation: Oak Ridge National Laboratory, Georgia Institute of Technology
Abstract Preview: Purpose: The fundamental concept of radiopharmaceutical therapy involves binding radioisotopes to molecules with a known biokinetic distribution, which ensures that the targeted organs receive higher ...
Authors: Julien Bancheri, Chau Giang Bui, David G Cooke, Christopher M Lund, Morgan J Maher, Jan P. Seuntjens, Jason Z Yuan
Affiliation: Princess Margaret Cancer Centre & University of Toronto, Medical Biophysics, University of Toronto, University of Toronto, Medical Physics Unit, McGill University, Department of Physics, McGill University
Abstract Preview: Purpose: The dielectric wall accelerator (DWA) offers a low-cost solution for proton therapy, using non-resonant waveguides to accelerate particles. Radial waveguides (RWGs), consisting of a dielectri...
Authors: Jadon O Buller, Yankun Lang, Liangzhong Xiang, Andrea Paulina Yanez
Affiliation: University of Maryland School of Medicine, University of California, Irvine
Abstract Preview: Purpose: With the widespread adoption of electroporation, there is growing interest in deep-tissue electric field monitoring. Electroporation therapy ablates target regions by generating electric fiel...
Authors: Piotr Pater, Nicholas G. Zacharopoulos
Affiliation: McGill University, Medical Physics Unit, Gerald Bronfman Department of Oncology, McGill University
Abstract Preview: Purpose: To develop a method for accurate relative film dosimetry that incorporates non-uniformity
corrections.
Methods: The accuracy of Devic’s [1] linearization approach was evaluated using si...
Authors: Tatsuya Fujisaki, Hiraku Fuse, Shin Miyakawa, Hiroki Nosaka, Masato Takahashi, Kenji Yasue
Affiliation: Ibaraki Prefectural University of Health Sciences
Abstract Preview: Purpose: This study aimed to evaluate the robustness of Jacobian-based and HUchange-based CT ventilation imaging (CTVI) against respiratory amplitude variations using a custom-designed alveolar phanto...
Authors: Yijian Cao, Sean T Grace, Marissa Joyce Vaccarelli
Affiliation: Northwell
Abstract Preview: Purpose:
This study investigates the dosimetric efficacy of significantly limiting gantry angles in prostate radiation therapy to as a promising strategy to maintain plan quality while enhancing ma...
Authors: Jingyao Chen, Yingli Yang, Jie Zhang
Affiliation: Department of Radiology, Ruijin Hospital, Shanghai Ruijin Hospital, Shanghai United imaging Healthcare Advanced Technology Research Institute
Abstract Preview: Purpose: Magnetic Resonance Electrical Properties Tomography (MR EPT) is a method to spatially mapping the conductivity and permittivity based on small B1 field changes after the imaged object was int...
Authors: Asma Amjad, Slade J. Klawikowski, Natalya V. Morrow, Haidy G. Nasief, Eric S. Paulson, An Tai, Hualiang Zhong
Affiliation: Department of Radiation Oncology, Medical College of Wisconsin
Abstract Preview: Purpose: Accurate and precise linac-based SRS commissioning can be very challenging. Thus, it is important to increase the confidence in the measurement at each step prior to end-to-end testing. The p...
Authors: Huixiao Chen, Zhe (Jay) Chen, MinYoung Lee, Sameer Taneja
Affiliation: Yale School of Medicine, Department of Therapeutic Radiology, Yale University School of Medicine
Abstract Preview: Purpose: Superficial radiation therapy uses low-energy x-rays to treat various types of cancers, including non-melanoma skin cancers, and dermatological conditions such as keloid scars, mycosis fungoi...
Authors: Ted Lynch
Affiliation: Sun Nuclear, a Mirion Medical Company
Abstract Preview: Purpose: A novel ultrasound quality assurance method is presented that constructs “synthetic spheres” at arbitrary locations within the image frame by combining correlation length measurements from a ...
Authors: Chonthicha Bunchuailuea, David Geeler, Martin Sabel, Claas Wessels
Affiliation: Varian Medical Systems
Abstract Preview: Purpose: Elongated treatment sites, such as craniospinal treatments, pose multiple challenges in today’s practices, requiring multiple CBCTs for position verifications per fraction, which might increa...
Authors: Les Butler, Joyoni Dey, Conner B Dooley, Victoria L Fontenot, Kyungmin Ham, Hunter Cole Meyer
Affiliation: Louisiana State University, Center for Advanced Microstructures and Devices, Linfield University
Abstract Preview: Purpose: X-ray interferometry is an emerging imaging modality with a wide variety of potential clinical applications, including lung and breast imaging. Small inaccuracies in grating position lead to ...
Authors: Julia Brosch-Lenz, Munir Ghesani, Francesc Massanes, Michael Morris, Babak Saboury, Eliot Siegel, Alexander Hans Vija
Affiliation: Institute of Nuclear Medicine, Siemens Medical Solutions USA Inc., Molecular Imaging
Abstract Preview: Purpose: Quantitative imaging of the radiopharmaceutical distribution is crucial for treatment evaluation and dosimetry. However, as patient numbers for Lutetium-177-(177Lu)-labelled therapies continu...
Authors: Jiankui Yuan, Dandan Zheng, Tingliang Zhuang
Affiliation: Medical Artificial Intelligence and Automation (MAIA) Laboratory, Department of Radiation Oncology, UT Southwestern Medical Center, University of Rochester, Varian Medical Systems, Advanced Oncology Solutions
Abstract Preview: Purpose: In Monte Carlo (MC) radiation therapy dose calculations, latent variance exists when directly applying phase-space files (PSF) with a finite number of source particles, while the latter is pr...
Authors: Go Akamatsu, Hiroaki Ikawa, Yuma Iwao, Han Gyu Kang, Masashi Koto, Hideyuki Mizuno, Akram Mohammadi, Makoto Shinoto, Minoru Tajiri, Hideaki Tashima, Chie Toramatsu, Taiga Yamaya
Affiliation: National Institutes for Quantum Science and Technology
Abstract Preview: Purpose: In-beam positron emission tomography (PET), in which positron emitters produced through fragmentation reactions are visualized, is a practical method for range verification in particle therap...
Authors: Laura I. Cervino, Linda X. Hong, Jessica Pagan, Angelica A. Perez-Andujar, Maria Thor
Affiliation: Department of Medical Physics, Memorial Sloan Kettering Cancer Center, Memorial Sloan Kettering Cancer Center, Memorial Sloan-Kettering Cancer Center
Abstract Preview: Purpose: Three-dimensional conformal radiotherapy (3DCRT) and Volumetric-Modulated Arc Therapy (VMAT) are used to treat locally advanced breast cancer. The 3DCRT treatment plans use the treatment fiel...
Authors: Steve B. Jiang, Ruiqi Li, Hua-Chieh Shao, Kenneth Westover, You Zhang, Tingliang Zhuang
Affiliation: Medical Artificial Intelligence and Automation (MAIA) Lab, Department of Radiation Oncology, UT Southwestern Medical Center, UT Southwestern Medical Center, Medical Artificial Intelligence and Automation (MAIA) Lab & Department of Radiation Oncology, UT Southwestern Medical Center, Medical Artificial Intelligence and Automation (MAIA) Laboratory, Department of Radiation Oncology, UT Southwestern Medical Center
Abstract Preview: Purpose:
Respiratory motion is a long-standing challenge for lung SBRT, particularly for centrally-located lung tumors where increased toxicity demands more precise motion management during treatme...